The Maximum Tolerated Dose of Gamma Radiation to the Optic Nerve during Gamma Knife Radiosurgery in an Animal Study

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Kunming Med Coll, Affiliated Hosp 1, Dept Neurosurg, Kunming, Peoples R China; [2]China Japan Friendship Hosp, Dept Neurosurg, Beijing 100029, Peoples R China; [3]China Japan Friendship Hosp, Dept Clin Lab, Beijing 100029, Peoples R China; [4]Huazhong Univ Sci & Technol, Union Hosp, Tongji Med Coll, Dept Neurosurg, Wuhan 430074, Peoples R China; [5]Huazhong Univ Sci & Technol, Union Hosp, Tongji Med Coll, Dept Ophthalmol, Wuhan 430074, Peoples R China
出处:
ISSN:

关键词: Gamma knife Optic nerve Radiation dose Electrophysiology

摘要:
Background: The safety of gamma knife radiosurgery should be considered when treating pituitary adenomas. Objectives: To determine the maximum tolerated dose of radiation delivered by gamma knife radiosurgery to optic nerves. Methods: An animal model designed to establish prolonged balloon compression of the optic chiasm and parasellar region was developed to mimic the optic nerve compression caused by pituitary adenomas. Twenty cats underwent surgery to place a balloon for compression effect and 20 cats in a sham operation group received microsurgery without any treatment. The effects of gamma knife irradiation at 10-13 Gy on normal (sham operation group) and compressed (optic nerve compression group) optic nerves were investigated by pattern visual evoked potential examination and histopathology. Results: Gamma knife radiosurgery at 10 Gy had almost no effect. At 11 Gy, P100 latency was significantly prolonged and P100 amplitude was significantly decreased in compressed optic nerves, but there was little change in the normal optic nerves. Doses of 11 Gy and higher induced significant electrophysiological variations and degeneration of the myelin sheath and axons in both normal and compressed optic nerves. Conclusions: Compressed optic nerves are more sensitive to gamma knife radiosurgery than normal optic nerves. The minimum dose of gamma knife radiosurgery that causes radiation injury in normal optic nerves is 12 Gy; however, the minimum dose is 11 Gy in compressed optic nerves. Copyright (c) 2013 S. Karger AG, Basel

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2014]版:
大类 | 4 区 医学
小类 | 3 区 外科 4 区 神经成像 4 区 神经科学
最新[2023]版:
大类 | 4 区 医学
小类 | 4 区 神经成像 4 区 神经科学 4 区 外科
JCR分区:
出版当年[2013]版:
Q2 SURGERY Q3 NEUROIMAGING Q4 NEUROSCIENCES
最新[2023]版:
Q2 SURGERY Q3 NEUROIMAGING Q4 NEUROSCIENCES

影响因子: 最新[2023版] 最新五年平均 出版当年[2013版] 出版当年五年平均 出版前一年[2012版] 出版后一年[2014版]

第一作者:
第一作者机构: [1]Kunming Med Coll, Affiliated Hosp 1, Dept Neurosurg, Kunming, Peoples R China;
通讯作者:
通讯机构: [2]China Japan Friendship Hosp, Dept Neurosurg, Beijing 100029, Peoples R China;
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:57189 今日访问量:0 总访问量:1788 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 昆明医科大学第一附属医院 技术支持:重庆聚合科技有限公司 地址:云南省昆明市西昌路295号(650032)