This study aims to evaluate the applications and values of dual-source computed tomography (DSCT) intelligent optimal kV scanning technology (CARE kV) in coronary CT angiography (CCTA). 150 patients with normal body mass index were performed DSCT coronary angiography, then randomly divided into the "Semi", 120,100 and 80 kV Group, and the 2 "on" groups, with 30 patients in each group. The first 5 groups used the reference voltage as 120 kV, and the reference current as 400 mAs, while the other group used the reference voltage as 100 kV, and the reference current as 400 mAs. The image quality, average CT value, image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and radiation dose were comparatively analyzed among the 5 groups. The image quality scores of the five groups showed no significant difference (P > 0.05); the average CT values and image noises had significance (P < 0.05), while SNR and CNR showed no significant difference (P > 0.05). The 80 kV group showed the biggest noise, with the CT value as 700 HU, while the radiation dose was the lowest, followed by the on group. As for the patients with normal body mass index (BMI), CARE kV-"on" could obtain high-quality images and lower radiation dose for CCTA, while the operation was simple and convenient.
基金:
Yunnan provincial Natural Science Foundation (2008c0477m), Project of Scientific Research Fund of Yunnan Provincial Department of Education (Project number: 2010Y177); Graduate Innovation Fund of Kunming Medical College (2009N22), multi-center study of MSCT coronary angiography (Project number: NCT01164839). Project of innovation team Fund of Yunnan provincial (2014HC018).
语种:
外文
PubmedID:
中科院(CAS)分区:
出版当年[2016]版:
大类|4 区医学
小类|4 区医学:研究与实验
最新[2023]版:
无
第一作者:
第一作者机构:[1]Department of Medical Imaging, The First Affiliated Hospital of Kunming Medical University, Kunming 650032, China
通讯作者:
通讯机构:[2]Department of Medical Imaging, Taizhou People’s Hospital, Taizhou 225300, China[*1]Department of Medical Imaging, The First Affiliated Hospital of Kunming Medical University, No. 295 Xichang Road, Kunming 650032, China.
推荐引用方式(GB/T 7714):
Zhang Jun,Kang Shaolei,Han Dan,et al.Application of intelligent optimal kV scanning technology (CARE kV) in dual-source computed tomography (DSCT) coronary angiography.[J].International journal of clinical and experimental medicine.2015,8(10):17644-53.
APA:
Zhang Jun,Kang Shaolei,Han Dan,Xie Xiaojie&Deng Yaming.(2015).Application of intelligent optimal kV scanning technology (CARE kV) in dual-source computed tomography (DSCT) coronary angiography..International journal of clinical and experimental medicine,8,(10)
MLA:
Zhang Jun,et al."Application of intelligent optimal kV scanning technology (CARE kV) in dual-source computed tomography (DSCT) coronary angiography.".International journal of clinical and experimental medicine 8..10(2015):17644-53